Setup Menu - DeFelsko PosiTector SHD Manual Del Usuario

Ocultar thumbs Ver también para PosiTector SHD:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

1 Pt Adjust
Adjusts the Gage to a known material hardness.
Step 1: Select 1 Pt Adjust from the Cal Settings menu.
Step 2: Press the (+) button to select the number of readings to be
used to obtain an average, typically 3 to 10 readings. The greater the
variation between readings, the more readings should be taken to
obtain an average.
Step 3: Repeatedly measure the known hardness sample using the
on-screen green arrows as a guide. After the final reading, the Gage
will calculate and display an average measurement value of all
readings taken on the sample. If the expected hardness value is not
obtained (within tolerance), adjust the displayed value up (+) or down
(-) to the known hardness value and press
Cal Reset
Restores the gage back to factory calibration and range settings.
The
icon will appear on the display.
Reset
Reset (menu reset) restores factory settings and returns the Gage
to a known condition. The following occurs:
- All batches, stored measurements, batch names and screen
captures are erased.
- Calibration adjustments are returned to factory settings.
- Menu settings are returned to the following:
Memory = OFF
Auto Sub-Batch = OFF
Statistics = OFF
Hi Lo Alarm = OFF
Hi Res = OFF
Cal Settings Menu

Setup Menu

4
to enter the value.
Bluetooth & Stream = OFF
WiFi & Access Point = OFF
Auto Ignore 20/90 = OFF
Auto Dim = ON
Cont. Reading = OFF