Cal Memory
When measuring a specific part with a specific powder, it is often
convenient to be able to store any unique calibration adjustments
made for that application. Then, if you return to that part, the
corresponding Cal setting can be conveniently and quickly
restored.
Cal Lock
When selected, the
are "locked" to prevent further user adjustments. Uncheck to make
further adjustments.
Reset
Reset (soft reset) restores factory settings and returns the Gage
to a known condition. The following occurs:
- All batches, stored measurements, images, and batch names
are erased.
- Menu settings are returned to the following:
Memory = OFF
Statistics = OFF
Perform a more thorough Hard Reset by powering down the Gage,
waiting several seconds, then simultaneously holding both the
center
This returns the Gage to a known, "out-of-the-box" condition.
It performs the same function as a menu Reset with the addition of:
- All calibration adjustments and Cal Memory are cleared and
returned to the Gage's factory calibration settings.
- Bluetooth Pairing info is cleared.
- Menu settings are returned to the following:
Units = microns
Flip Display = Normal
Auto Sync = OFF
- A probe Zero (pg 6) must be performed after a Reset.
NOTES:
- Date, Time and WiFi are not affected by either Reset.
✔
icon will appear and all calibration settings
Setup Menu
Setup Menu
Bluetooth = OFF
Display = None
and (+) buttons until the Reset symbol
White on Black = OFF
Language = English
Battery Type = Alkaline
9
Cal Lock = OFF
appears.
Backlight = Normal
USB Drive= ON