Weighing; Performing A Weighing Operation - Seca 745 Instrucciones De Usuario

Ocultar thumbs Ver también para 745:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 91
8.

5.2 Weighing

Performing a weighing
operation
38 •
5. Check whether the beam is balanced.
Pointer and counterpointer must be at exactly or
approximately the same level.
6. Repeat Step 3 if required.
7. Fix the tare weight in position using the knurled
screw.
8. Turn the fine-tare screw until the pointer and
counterpointer are exactly opposite one another.
The scale is operational.
7.
CAUTION!
Risk of falling, faulty operation
To prevent falls from the scale and faulty opera-
tion, you must ensure before each use that the
scale is operational.
– Ensure that the tray is correctly fixed in
– Perform a function check (see "Function
– Check whether the beam is balanced (see
1. Ensure that the locking lever of the beam is re-
leased (see "Releasing/locking the beam" on
page 36).
2. Place the baby on the tray.
position (see "Put on tray" on page 35).
check" on page 41).
"Checking the position of the beam" on
page 36).

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

725

Tabla de contenido