2 Pt Adjust
Provides greater accuracy within a limited, defined range. Measure
a thin value (often zero) and a thicker value. These values should
bracket your expected thickness range.
Cal Lock
When checked, the
are "locked" to prevent further user adjustments.
N Lock
Use when operating regularly on non-ferrous substrates. The
icon appears and the probe will only use the eddy current principle
to shorten measurement time and extend battery life. Useful when
measuring coatings over plated steel.
Cal Memory
The current cal setting (ex. Cal 3) displays in the upper right corner
of the LCD. A stored cal setting can be restored when you return
to a particular part. Cal 1 can be adjusted but never deleted, and
is always made active with factory settings after a Reset (pg. 4). A
new cal setting cannot be created if a batch is open with readings.
Reset
eset (soft reset) restores factory settings and returns the
instrument to a known condition. The following occurs:
- All batches, stored measurements, images, and batch names
are erased.
- All calibration adjustments and Cal Memory are cleared and
returned to the Gage's factory calibration settings.
- Menu settings are returned to the following:
Memory = OFF
High Res = OFF
Statistics Mode = OFF
Hi Lo Alarm = OFF
Scan Mode = OFF
icon appears and the current Cal settings
(FN ferrous/ non-ferrous cobmination probes only)
(Advanced models only)
Setup Menu
Setup Menu
4
Cal Lock = OFF
Bluetooth = OFF
Display = None
N Lock = OFF