Product Description - GCE MEDIMETER Instrucciones De Uso

Ocultar thumbs Ver también para MEDIMETER:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 51
OPERATING CONDITIONS
EN
In case of storage at a temperature below -20 °C, do not operate the flow
selector until it has been allowed to increase its temperature to a minimum
of -20 °C.
4. PERSONNEL INSTRUCTIONS
The Medical Devices Directive 93/42/EEC states that product provider
must ensure that all personnel handling the product are provided with the
operating instructions & performance data.
Do not use the product without properly familiarization of the product
and its safe operation as defined in this Instruction for use. Ensure user
is aware of particular information and knowledge required for the gas
in use.

5. PRODUCT DESCRIPTION

Gas from the medical gas pipeline system is entering through the inlet con-
nector (A). Flow is then adjusted to the request-
ed value by turning handwheel (B), which con-
trols the valve for flow regulation. The adjusted
flow is visible on the calibrated tube via the float
(C). Gas flows through the tube, when constant
flow is adjusted, the float is stabilized in a cer-
tain position (a certain flow on the scale). Gas
is delivered to the user through the flow outlet
user connection (D).
A. INLET QUICK CONNECTOR
The flowmeter is fitted to the medical gas pipe-
line system terminal unit or to medical regulator
by a gas specific male probe for specific gas.
There are two basic variants:
1 – inlet connector is directly mounted on the
flowmeter,
2 – connection to the rail, including hose and
inlet connector.
-20/+60 °C
10/100 %
600/1200 mbar
STORAGE AND TRANSPORT
CONDITIONS
-30/+60 °C
10/100 %
600/1200 mbar
A
D
4/124
C
B

Publicidad

Tabla de contenido
loading

Tabla de contenido