5.2.1.5. Zero adjustment.

After disconnecting the voltage and current input channels from the measuring grid, place the cursor on
START and press
be displayed for the duration of the zero adjustment procedure (10-20"). When the procedure is complete,
the system will automatically return to the CONNECTIONS SETUP page.

5.2.1.6. Connection check.

Once the instrument has been configured and connected to the system, the instrument can check if the
connection to the electrical system has been performed correctly (to perform this check, the PF value must
comply with the value indicated on the screen).
Place the cursor on Check and press
Select "Repeat" to perform a new check.
Select "Exit" to return to the CONNECTIONS SETUP page.

5.2.2. Current probes setup.

Due to automatic recognition of current probes, the setup values will be those detected at power up. If you
need to use different clamps from those recognized in power on, you will have to manually change the setup
as shown below, or alternatively, make a new power on after connecting the new probes.
to correct the offset, in case the latter has deviated. A page with numerical values will
to perform the check. The related outcome will then be displayed.
Voltage phase sequence
Threshold of the measured PF which allows for a correct analysis
(if the PF is lower than this value, the check cannot provide valid
information)
Check of the correspondence between voltage and current of
each phase and posible error message:
This page allows the users to select:
the type of probe used for I1, I2, I3, selecting aming Flex (non-
amplified flexible sensors) or AC/DC (clamp);
the sensor transformation ratio on I1, I2, I3 (press and hold down
or
the type of probe used for In, selecting among Flex (non-
amplified flexible sensors) or AC/DC (clamp);
the sensor transformation ratio on In (press and hold down
to increase scrolling speed);
the type of probe used for Iaux selecting among Flex (non-
amplified flexible sensors) or AC/DC (clamp);
the sensor transformation ratio on Iaux (press and hold down
or
Ok = Connection is correct
Invertir CT = Invert the direction of the current clamp
indicated
Fallada = No correspondence between voltage and
current or the PF value is lower than the threshold displayed.
to increase scrolling speed);
to increase scrolling speed);
- 67 -
- 67 -
or