Maintenance
Fault/message
Volume too great
Sample interruption
Fault number
6
Maintenance
Have the device calibrated every 12 months, as drift effects increase the
uncertainty of measurement over time. Calibrationand maintenance should only
be carried out by authorized persons, e.g., DrägerService.
The last 30 days before next adjustment / service is needed, a day count down
will be shown as information in the display after switching on the device. You
have to confirm this information with the
date the device starts to count up how many days it is over requested adjustment
/ service. This information has to be confirmed by pressing the
It shall remember you that the device needs to be adjusted / serviced.
Have the instrument re-calibrated, if an accuracy check does not produce a
result per NHTSA's EBT Model Specifications. Calibrations should only be
carried out by authorized persons, e.g. DrägerService
6.1

Replace battery

1.
Switch off the device.
2.
Slide the battery cover off.
3.
Replace the battery (battery type: 3 V CR123A; order no.: 83 21 390),
paying attention to the correct polarity.
4.
Fit the battery cover.
5.
Check that the time is correctly set.
14
Cause
Measuring range
exceeded.
Irregular exhalation,
abrupt end, or inhalation
Instrument fault
Remedy
Repeat breath sample.
Repeat breath sample.
Switch the instrument off
and on again. If the fault
remains, contact
DrägerService.
button. After the adjustment / service
button as well.