www.allice.de
R&S ENV216
2 Brief introduction
2.1 Introduction
Operating Manual 5201.6693.14 - 03.00
The R&S ENV216 compact Two Line V-Network is used to measure noise voltage on
mains-dependent consumers.
Its principal tasks are:
●
To supply the equipment under test with mains voltage
●
To provide a standardized load impedance
●
Defined transmission of the noise voltage generated by the equipment under test
to the EMI test receiver
●
Isolation of the test circuit from interference of the power source.
The R&S ENV216 Two Line V Network is constructed using air-core inductances
(50 µH and 250 µH) and conforms with the recommendations of VDE0876 and
CISPR 16-1-2 / 16-2-1.
Other features of the equipment include:
●
Artificial Hand
●
10-dB-attenuator in the RF path
●
Switchable 150-kHz-highpass filter
●
Switchable Pulse voltage limiter at the measurement output
●
Remote control port
●
External power supply via plugged power pack for tests on non-standardized a.c
and d.c. voltages
The network may be operated manually or by remote control with TTL levels.
Allice Messtechnik GmbH
Brief introduction
4