3.3 Protective grounding
Operating Manual 5201.6693.14 - 03.00
Improper or negligent handling could result in death!
Operating the device without protective earth may cause serious injury or death.
The direction in chapter 1 entitled "Safety direction" must be followed without
exception!
Improper or negligent handling could result in death!
Before using for the first time, the network must be connected to an additional
conductor that is compliant with VDE0100.
Users must be aware that it is possible for socket connections and grounding
conductors to become disconnected. A further grounding conductor with adequate
diameter must be connected between a grounding conductor connection for the
measurement area and the grounding bolt (1) on the R&S ENV216.
Only then may the network be connected to the mains power supply.
At shutdown, the same procedure must be followed in the reverse order:
First the R&S ENV216 must be disconnected from the mains power supply, and only
then should the additional grounding conductor connection be removed.