3.4 Mains connection
Operating Manual 5201.6693.14 - 03.00
Connection to the mains power supply is provided by a permanently attached mains
cable.
The device is not equipped with an ON / OFF switch. Instead, the device plug of the
connecting cable serves for disconnection from the AC supply (for this see also the
corresponding section in the basic safety instructions).
Improper or negligent handling could result in death!
When beginning operation, the sequence below must be followed strictly:
Connect network to auxiliary protective grounding
Connect mains plug to the mains power supply
Switching off:
Unplug network from the mains power supply
Disconnect auxiliary grounding connection
Improper or negligent handling could result in death!
Due to the high level of earth leakage current, the device cannot be operated on
power supplies protected by residual current devices (RCDs).
The power source must be protected by a circuit breaker with a maximum rating of
16 A characteristic B or C.
If the phase connection is made correctly, LED L1 (1) must light up.
If it does not:
●
For mains plugs with symmetrical grounding connection (e.g. Schuko plugs) the
plug's polarity must be reversed.
●
For mains plugs with asymmetrical grounding connection, phase L1 and neutral N
must be swapped by reconnecting.
●
For country- specific sockets see appendix A.