R&S ENV216

5 Remote control

Operating Manual 5201.6693.14 - 03.00
The functions L1, N and 150 kHz HP can be controlled remotely by applying a static
LOW level to the appropriate contact of the SUB-D25 female connector (1) on the rear
of the R&S ENV216.
1
As soon as a LOW signal is detected at input L1, N or 150 kHz HP, the corresponding
function is activated and manual operation is blocked at the same time. The functions
that have been switched on are indicated by the associated LEDs. When remote
control is active, the REMOTE LED is lit.
When the LOW signal is removed, manual operation is enabled again and the status
set manually prior to the remote operation is restored.
The settings are not saved if the device is disconnected from mains supply; instead,
the default status ( L1 "ON" and 150 kHz HP "OFF") is always restored when mains
power is switched on.
Remote control device
The R&S ENV216 is configured for direct connection to ROHDE&SCHWARZ
test receivers.
If remote control is performed via an R&S receiver, the REMOTE LED is lit when the
ESH3-Z5 type network is selected. The user can choose between "N" and "L1".
The 150-kHz-highpass filter can be inserted into the signal path by pressing the
"Floating" key on the receiver.
Allice Messtechnik GmbH
CAUTION
(High leakage current of approx. 800 mA)
Put unit into operation in the following order:
1. connect artificial mains network to protective earth
using additional connector
2. Insert power connector in socket of supplying network
Put unit out of operation in the following order:
1. Disconnect artificial mains network from supplying network
2. Disconnect additional protective earthing
REMOTE CONTROL
EXT POWER
CONNECT TO MAINS SU
SUPPLY
WITH PROTECTIVE EAR
Remote control
(Ho
Inbetriebnahme in fo
1. Netznachbildung
2. Netzstecker mit
Außerbetriebnahme:
1. Netznachbildung
2. zusätzliche Schu
26