2. Safety Information

The security information included in this chapter should be
read carefully.
The user should check for compliance with security conditions
set out in this Manual to ensure the safe use of the equipment.
In case of doubt, contact the Technical Assistance Service.
Safety information is related to both user protection and
equipment protection.
Access to the equipment interior
Only authorised technical service may access the equipment
interior to carry out maintenance and repair tasks other than
the routine replacement of expendable items.
Electrical connections
-
Check that connection to the power system is 230
VAC, 50 -60 Hz.
-
Only use the transformer supplied with the
equipment.
-
Supply voltage fluctuations should not exceed 10%
of the rated voltage.
-
Supported transitional overvoltage: The typical
present in the power supply network (category II of
the standard IEC 60634-4-443).
-
The electrical plug should not be situated below
the equipment.
-
The electrical plug should be accessible in case of
be necessary to disconnect the equipment quickly.
Explosive atmospheres
The equipment is not designed to work in explosive
atmospheres.
Environmental pollution permitted
Pollution grade 1 (No environmental pollution or only dry
pollution, not conductive).
Feed water and hydraulic connections
The feed water must comply with the specifications listed in
this Manual.
Feeding poor quality water (turbidity and/or high hardness,
inadequate conductivity, unwanted pollutants, etc.) can result
in deterioration of the equipment components or a lower
performance.
Respect the pressure conditions established for the feed water
to the equipment. High pressure can damage the hydraulic
system. In these cases, it is necessary to install a pressure
regulator before the equipment.
A feed water pressure lower than the minimum required will
cause poorer performance and improper functioning of some
of the system components.
Have a stopcock that is easily accessible when you need to
cut the water supply to the equipment.
Locate the equipment so that the electric and
hydraulic connections located on the rear of the equipment are
not adversely affected.
Prevent straining the power cable or hydraulic connections.
Expendable items
Replace the expendable items in the times anticipated and
based on the messages from your equipment.
Use only original expendable items.
components may cause equipment damage and loss of
guarantee.
To move the equipment, handle it by holding at the
base.
Using non original