You can reset your device to factory default. This will delete the entire memory and
the settings for language, date and time will be reset.
•
Turn of the device.
•
Switch on the device with the
logo with revision stand), simultaneously press the
•
As soon as the screen turns orange, release the buttons and press the
-icon again.
•
You are now in the "Reset" menu. By pressing the
reset your device to the factory settings.
•
Switch on the device and carry out the initial configuration
(as described in chapter 1.3.4).
5
NOTES
5.1
PATIENT SAFETY
Indication: electronic blood pressure monitor to measure blood pressure using
inflatable cuffs. It is intended for preventive use, for regular blood pressure checks
performed by a doctor or qualified medical staff.
The EN200 BP is not intended for use with the following groups of people:
•
No measurement for newborns, toddlers and children.
•
Not suitable for people with relative and absolute contraindicators.
•
Not suitable for patients with pre-eclampsia.
•
Not suitable on the arm on the side of a mastectomy.
Relative contraindicators:
• Lymphoedema
• Partial or complete paralysis
• Arterial or venous vascular accesses (e.g. Viggo)
Setting options:
• Date
• Time
• Language selection
Proceed as described in 1.3.4
MED 113275
2017-05-02
-button. During the start screen (HEINE
and
keys.
-symbol, you will
57/176