•
To ensure all-pole disconnection from the mains at all times, the
ME device must be installed so that the switch mode power supply
is accessible and disconnectable.
•
If liquid has been sprayed onto the device, send the device to your
authorised HEINE sales partner or directly to the manufacturer for
checks.
•
Comply with the statutory provisions and the current state-of-theart
when cleaning and disinfecting the device. Use the detergent and
disinfectant recommended by the manufacturer to clean and disinfect
the device. Clean and disinfect your device and accessories in line with
the specifications in Chapter
and 5.4. Cleaning and maintaining the cuff.
5.2
OPERATION, MAINTENANCE AND CARE
•
Metrological control (MTK): the device is calibrated by the manufacturer
for a period of two years. The metrological control must take place in
accordance with §11 MPBetreibV (Medical Device Operator Regulation)
at least every two years and after repairs. The inspection is subject to
a charge and may be performed by the manufacturer or an authorised
maintenance service centre, in accordance with the Medical Device
Operator Regulation. To perform the MTK, the "MTK Calibration Procedure"
document must be requested from your authorized HEINE sales partner or
the HEINE Customer Service.
•
A regular, mandatory safety check (STK), at least every two years, is
prescribed for this device pursuant to the MPBetreibV. In addition to the
statutory requirements, the manufacturer prescribes a regular annual
safety check; this must also be performed by an authorized service centre
after every repair, change or modification of the system or device.
Every safety check must be completed in full and consists of the following
components: visual inspection, functional check, testing of the monitoring,
safety, display and messaging systems, measurement of the key safety
values, electrical test.
•
Only use the device in accordance with its intended use defined in the
instructions for use. The warranty expires if the device is used incorrectly.
•
The compression or other mechanical restriction of the crosssection of
connecting tubes must be avoided.
60/176
5.3 Cleaning and maintaining the device
MED 113275
2017-05-02