4 GETTING STARTED (continued)
4.6 SELECTING THE MEASUREMENT UNITS
The Elcometer 311 can take measurements in µm and mils. To set the
measurement units:
1
Press the '
é
2
Use the
alternate between 'µm' and 'mil', followed by the ç softkey to
return to the main reading screen.
5 TAKING A READING
Before taking a reading, it is good practice to check the performance
of the gauge using the calibration check procedure, see Section 7
'Checking the Gauge Calibration' on page en-6.
To take a reading:
1
Whilst holding the gauge perpendicular, gently place the probe
onto the surface.
The reading may be inaccurate if the probe is not held flat against the
4
surface.
2
The coating thickness is shown on the display.
'>500µm' ('>20mils') indicates a reading outside the gauge range.
4
3
For subsequent readings, lift the probe off and then replace it on
to the coated surface.
Using ElcoMaster the Elcometer 311 Model T can simultaneously
transmit the live reading to a PC or mobile device for report generation,
see Section 8 'Downloading Data' on page en-7 for more information.
The Elcometer 311 Model T calculates and displays statistical values
as each reading is taken, see Section 6 'Statistics' on page en-6 for
details.