9 TAKING A READING

9.1 BEFORE YOU START

1
Press the On/Off button to switch the gauge on.
2
Connect a transducer to the gauge.
All single element delay line transducers which can be connected directly to
the base of a PTG gauge - see Section 16.1 'Transducers' on page en-28 -
are 'intelligent' transducers and will be identified automatically by the gauge.
If using other Elcometer 'non-intelligent' single element delay line
transducers or other manufacturers' transducers, a transducer adaptor is
required - see Section 16.5 'Transducer Adaptor' on page en-31.
3
Select the measurement mode - see Section 4.6 on page en-8.
If using Plastic Mode (PLAS), a graphite delay line must be fitted to the
transducer as standard acrylic delay lines (supplied with each transducer)
are not suitable. Graphite delay lines are available to purchase as
optional accessories - see Section 16.2 'Delay Lines' on page en-29.
4
Calibrate the gauge - see Section 7 on page en-13.
5
Prepare the test surface - see Appendix 1 on page en-34.
9.2 TAKING A READING IN STANDARD MODE
1
Apply a small amount of couplant to the test surface.
2
Press the transducer into the couplant, making sure it is flat
against the surface.
Moderate pressure on the top of the transducer using the thumb or index
finger is sufficient; it is only necessary to keep the transducer stationary
and seated flat against the surface of the material.
3
The display will show a value which is constantly updating.
The gauge will take 4, 8 or 16 readings per second as selected
by the user via Menu/Setup/Reading/Reading Rate.
The stability of the reading is indicated on the stability bar to the right of
the display. A valid reading has a stability of 5 or more. If the stability
indicator has fewer than 5 bars showing or the numbers on the display
seem erratic, make sure there is an adequate film of couplant beneath the
transducer, and that the transducer is seated flat against the material. If
the condition persists, it may be necessary to select a different transducer
(size o frequency) for the material being measured.
r
4
PTG6 - The last reading is held on screen when the transducer
is removed from the surface.
PTG8 - Press 'Save' to store the current reading in the gauge or
batch memory and remove the transducer from the surface.