Cleaning And Disinfection - Gima 28066 Guía De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 41
4. Install medication cup to main unit.
5. Assemble the mask, as shown to the side:
Operation method:
1. Turn on the power: Press "ON/OFF" button for more than 1 sec-
ond, power indicator (green) lights and device starts atomizing.
Attention:
If medication cup is not loaded with any medication, the de-
vice will automatically shutdown after power Indicator lights up
about 1s.
After startup, the quantity of medication mist may change at the
beginning of device working, which Is a normal phenomenon.
2. Inhalation: Hold the device in hand, put on the mask, slowly
inhale the medication mist.
Attention:
1) The angle of inclination should be within 45° during nebulizing.
2) During use, please do not strongly shake the device to prevent abnormal use.
3) Duration of each inhalation should be no more than 20 minutes.
4) Nebulizing treatment Is easy and comfortable, if you feel uncomfortable during use, please stop the treatment.
3. Turn off the power. When the treatment is over, and medication almost runs out, the nebulizing sheet will generate a high-fre-
quency sound, and then the device automatically turns off. If you need to shut the device down during use, please press the
"ON/OFF" button for more than 1 second.
Attention:
At the end of treatment, it is normal that a little medication remains left in medication cup after automatic shutdown.
Chapter 5
MAINTENANCE, TRANSPORT AND STORAGE

5.1 Cleaning and disinfection

Clean and disinfect the device after each use. If the device is not cleaned, the nebulizing will be affected because of drying and
coagulation of medication.
1. Remove medication cup, accessory and batteries from the main unit.
2. Open the cup lid and discard residual medication.
3. Add 75% ethanol solution in medication cup, cover the cup lid then leave for at least 10mins; it is advisable to gently shake it
for better disinfection.
4. Immerse the accessories to be disinfected into a container with ethanol solution, and put a lid on the container. Use 75%
ethanol solution soaking for 10mins or longer.
5. Discard the disinfectant in medication cup, take accessories out from the disinfectant; clean the medication cup and accesso-
ries with clear-water repeatedly.
6. Fill medication cup with clear-water, assemble it to the main unit, let the device work 10mins in order to clean the nebulizing
sheet.
7. After cleaning, use new medical gauze to wipe away the water, and fully dry,
8. Use 75% medicinal alcohol to wipe the surface of main unit, then air-dry or wipe-dry with a clean, soft cloth.
9. After all steps above, store the main unit, medication cup and accessories in a dry, clean place.
Attention:
Do not throw medication cup and accessories into boiling water for disinfection, otherwise the part may be deformed. Do not
put them in a microwave oven for drying.
The parts disinfected with disinfectant must be fully cleaned, or the residual disinfectant may cause symptomatic deterioration.
5.2 Medication cup replacement
The nebulizing sheet is a kind of consumable. In general, the service life of the atomizer is about six months (20 minutes per
time, three times a day).
Its service life depends on the use, medication, and the degree of cleaning. If no atomizing or little atomizing appears when the
device is working, please replace medication cup in time. (If you need to purchase medication cup, please contact the dealer.)
5.3 Transport and storage
Transport and storage environment:
Temperature: -40°C~+55°C
Relative humidity: 5%~96%
Atmospheric pressure: 500 hPa ~ 1060 hPa