Battery Charge Level

The LED Charge Status Indicator displays the charge level of each battery.
The charge level may be checked at any time by pressing the TEST button
and reading the number of illuminated LEDs. Each LED indicates
approximately 20% of the charge level. All five LEDs will illuminate when the
Battery has 81 – 100% power capacity. (See Table 4.2 for more information.)
Charge Status
Indicator
After a battery is inserted into the Power Manager, wait a minimum
of five (5) seconds for the Power Manager to verify an approved
power source is present. After this automatic verification, the LED
display on each battery will illuminate reflecting the current level of
charge. When a battery is being recharged, the LED indicating the
current charge level will flash approximately one time per second
(1 Hz).
When a battery is fully charged (100% capacity), all five LEDs will
illuminate for a brief period, and then extinguish.
st
The 1
(2 Hz), without pressing the TEST button, when the battery has
10% or less capacity while inserted in the SprintPack Power
Manager.
When two SprintPack Batteries are present, the battery with a lower
charge level will be used first. When the first battery becomes
depleted, the SprintPack Power Manager automatically switches to
the second battery to maintain a constant power output until both
batteries are depleted.
32328-001 Version A
(yellow) LED will flash approximately two times per second
Operator's Manual – SprintPack
TEST
Button
4-3