Symptoms
Continued...
SprintPack Battery is not
charging.
SprintPack Battery
Charge Status
Indicator LEDs do not
light up.
SprintPack Battery not
lasting 2.5 hours.
32328-001 Version A
Possible Causes
Battery was not
discharged below 80%
(4 lit LEDs), and was
not removed from the
Power Manager after
the last use.
Battery charge levels
are more than 40%
apart.
SprintPack Li-Ion
Power System error.
SprintPack Batteries
are depleted.
SprintPack Li-Ion
System needs to be
reset.
Battery status indicator
and charge level are out
of calibration.
Operator's Manual – SprintPack
What To Do
Remove battery from
Power Manager and then
reinsert.
Battery with lower charge
level will charge until it is
within 40% of the second
battery. Once the
batteries are within 40%
of each other both
batteries will charge
simultaneously.
Remove power source,
wait a minimum of 10
seconds, and reconnect
power source to reset the
SprintPack Li-Ion Power
System.
Recharge the batteries
using an approved power
source (see Table 4.1).
Remove power source,
wait a minimum of 10
seconds, and reconnect
power source to reset the
SprintPack Li-Ion Power
System.
Fully discharge and
recharge the battery.
™
5-5