Troubleshooting

This section describes troubleshooting for the SprintPack Li-Ion System. The
troubleshooting tables are organized by symptom. Contact Vyaire Medical
for information about symptoms that are not listed here.
Symptoms
The LTV Ventilator
EXTERNAL POWER
LED is not illuminated.
All five LEDs are blinking
on the SprintPack
Battery.
SprintPack Battery is not
charging.
5-4
Possible Causes
External power cords
loose.
SprintPack Batteries are
depleted.
SprintPack Li-Ion
System needs to be
reset.
System error (possible
causes include operating
temperature out of
range, temperature
mismatch between
Power Manager and
battery, or no valid
power sources).
Operating temperature
out of range.
Operator's Manual – SprintPack
What To Do
Disconnect and
reconnect the cords from
both the Power Manager
and the ventilator
connectors. Verify that
all connectors are free of
debris and securely
locked into their proper
outlet.
Check the Charge
Status Indicator on each
battery to verify charge
level.
Recharge the batteries
using an approved
power source (see Table
4.1).
Remove power source,
wait a minimum of 10
seconds, and reconnect
power source to reset
the SprintPack Li-Ion
Power System.
Allow the battery to
return within operating
temperature (see page
8-1) or connect a valid
power source and wait a
minimum of 10 seconds.
Allow the battery and/or
Power Manager to return
to operating temperature
specified on page 8-1.