Serter life of use
The serter has a minimum service life of one year.
Serter maintenance
The serter is intended for single-patient use. The serter needs to be cleaned after every use or whenever
there is debris, blood, or other contaminants.
Cleaning
To clean the serter, you will need the following materials: mild liquid soap, a soft-bristled toddler
toothbrush, and a container.
To clean the serter:
1. Prepare a mild liquid soap solution using 5 milliliters (1 teaspoon) of mild liquid soap per 3.8 liters
(1 gallon, which is equal to 16 cups) of tap water.
2. Rinse the serter under running tap water at room temperature for at least one minute. Continue
rinsing until the serter is visibly clean. Make sure all hard-to-reach areas are rinsed completely.
3. When rinsing the serter, press and release the mechanism on the underside of the serter to
ensure that the entire device is completely rinsed with water.
4. Submerge the serter in the mild liquid soap solution and soak it for at least 10 minutes.
5. Brush the entire surface of the serter using a soft-bristled toddler toothbrush until visibly clean.
Pay close attention to brushing the hard-to-reach areas.
6. Rinse the serter under running tap water at room temperature for at least one minute, and until all
visible liquid soap is gone.
7. Shake off any excess water or moisture and place the serter upright on a clean, dry, cloth to air
dry.