A
B
C
30233040B PBK9- / K...(x-T4)-series
5 Technical data and operating limits
5.1
Maximum verfication scale interval
Verification scale interval [e]
Verification scale interval [e]
K...(x-T4)
3 x 6000 e
0.6 kg / 0.1 g
Multi Interval
1.2 kg / 0.2 g
max / e
3 kg / 0.5g
1 x 30000 e
Single Range
K...(x-T4)
A15
3 x 3000 e
3 kg / 1 g
Multi Interval
6 kg / 2 g
max / e
15 kg / 5 g
1 x 15000 e
Single Range
1 g
5.2 Maximum permissible load
All weighing platforms and the MPGI high precision load cell are equipped with an
overload protection. However, if the load exceeds the maximum permissible load, damage
of mechanical parts may occur.
The static load-bearing capacity, i.e. the maximum permissible load, is dependent on the
type of loading (positions A – C).
PBK98_-
Position
A central load
B side load
one-sided
C
corner load
PBK989-
XS0.6
A3
0.01 g
0.1 g
PBK98_-
AB60
B60
2 g
5 g
A3
A6
1.2 kg / 0.2 g
3 kg / 0.5 g
6 kg / 1 g
1 x 30000 e
0.1 g
0.2 g
A32
B60
6 kg / 1 g
15 kg / 5 g
12 kg / 2 g
30 kg / 10 g
32 kg / 5 g
60 kg / 20 g
1 x 32000 e
1 x 6000 e
1 g
10 g
A3
AB15
XS0.6
A6
AB30
AB60
2.0 kg
20 kg
50 kg
80 kg
1.5 kg
15 kg
40 kg
60 kg
1.0 kg
10 kg
30 kg
40 kg
A6
AB15
AB30
0.2 g
0.5 g
1 g
B120
CC150
CC300
10 g
10 g
20 g
CC150
CC300
30 kg / 10 g
60 kg / 20 g
60 kg / 20 g
150 kg / 50 g
150 kg / 50 g
300 kg / 100 g
1 x 15000 e
–
10g
CC150
B60
B120
CC300
150 kg
150 kg
500 kg
100 kg
100 kg
300 kg
50 kg
50 kg
150 kg
Technical data and operating limits
13