SuperSonic Imagine Aixplorer MultiWave Guía De Usuario De Fusión/Navegación página 42

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

How to Connect the Sensors to
the driveBAY
Connect the transducer sensor to the 'Sensors' connector marked with a
'1'. Rotate the sensor plug until it aligns with the holes on the 'Sensor'
connector, and then push until the plug clicks into place.
To connect the CIVCO devices, VirtuTRAX™, eTRAX™ and
omniTRAX™, plug them as follows: VirtuTRAX™ to the connector
marked '2', eTRAX™ to the connector marked '3', and omniTRAX to the
connector marked '4'.
CAUTION
Do not put the transmitter or the sensor near metallic objects.
CAUTION
Always connect the sensors and the transmitter to the driveBAY when
the Aixplorer® is off.
Turning On the Aixplorer®
Since the driveBAY is connected to the Aixplorer® power supply, the
driveBAY starts up when you turn on the Aixplorer®.
The indicator light (on the front of the drive BAY) slowly blinks with
a yellowish light while the driveBAY initializes. When complete, the
indicator light turns to a slow blinking green. This takes about 10 seconds.
NOTE
If a slow blinking green light does not appear, then the driveBAY has not
been properly initialized.