NOTE
You may delete all targets or risk areas at once by clicking on
All
.
Virtual probe model orientation
Free rotation
Set initial position
probe yourself
View from below
and could be used to verify the position of either the VirtuTRAX™ or
eTRAX™ devices
Selecting the Interventional Device
If they are correctly connected, click on either eTRAX™
On/Off
VirtuTRAX™
NOTE
There are 2 methods for the VirtuTRAX™ calibration in the Settings
Menu, under Other Settings tab:
• Manual method
A popup window will appear asking you to manually enter the needle
distance, which is the distance between the needle tip and the bottom
part of the needle holder
• Pivot method
Do a pivot movement with the VirtuTRAX™ needle holder by putting
the needle tip on a sterile and rigid plane surface.
Follow the on-screen instructions to perform the pivot calibration.
70
Aixplorer® Fusion/Navigation User's Guide
allows you to show the real orientation of the probe
allows you to set the initial position of the
allows you to view the probe model from below
to activate one of the devices.
Delete
On/Off