Hazardous (Classified) Location
Class I,Division2, Groups, A,B,C,D
OR
Class I, Zone 2, IIC
1
2
5350A
5350A
Approved
Termi-
6
3
nation
5
4
SENSOR
SENSOR
Terminal 3, 4, 5, 6
Vt or Uo : 5.71 V
It or Io : 8.4 mA
Pt or Po : 12 mW
Ca or Co : 40 F
La or Lo : 200 mH
Terminal 1.2
Ci: 2.0 nF
Li: 1 H
T1..T4
-40C Ta +85C
T5
-40C Ta +75C
T6
-40C Ta +60C
See installation notes:
Revision date:
Version /Revision
2009-09-29
V3/R0
5350V111
5350QE01
WWW.PRELECTRONICS.COM
Unclassified Location
32V
Class 2
Power Supply
5350A
This device must not be
connected to any
associated apparatus
which uses or generates
more than 250 VRMS
SENSOR
Page:
3/6
LERBAKKEN 10, 8410 RØNDE DENMARK.
Installation notes:
FM / CSA:
For installation in the US the 5350 shall be installed according to the National
Electrical Code (ANSI-NFPA 70).
For installation in Canada the transmitter shall be installed in a suitable enclosure to
meet installation codes stipulated in the Canadian Electrical Code (CEC).
The entity concept:
Equipment that is FM / CSA-approved for intrinsic safety may be connected to
barriers based on the ENTITY CONCEPT. This concept permits interconnection of
approved transmitters, meters and other devices in combinations which have not
been specifically examined by FM / CSA, provided that the agency's criteria are met.
The combination is intrinsically safe, if the entity concept is acceptable to the
authority having jurisdiction over the installation.
The entity concept criteria are as follows:
The intrinsically safe devices, other than barriers, must not be a source of power.
The maximum voltage Ui (V
) and current Ii (I
MAX
which the device can receive and remain intrinsically safe, must be equal to or
greater than the voltage (Uo or V
or V
OC
which can be delivered by the barrier.
The sum of the maximum unprotected capacitance (C
and the interconnecting wiring must be less than the capacitance (C
safely connected to the barrier.
The sum of the maximum unprotected inductance (L
the interconnecting wiring must be less than the inductance (L
connected to the barrier.
The entity parameters Uo,V
or V
and Io,I
OC
t
provided by the barrier manufacturer.
FISCO/FNICO rules:
The FISCO Concept allows the interconnection of intrinsically safe apparatus to
associated apparatus not specifically examined in such combination. The criterion for
such interconnection is that the voltage (Vmax), the current (Imax) and the power (Pi)
which intrinsically safe apparatus can receive and remain intrinsically safe,
considering faults, must be equal or greater than the voltage (Uo, Voc, Vt), the
current (Io, Isc, It,) and the power (Po) which can be provided by the associated
apparatus (supply unit). In addition, the maximum unprotected residual capacitance
(Ci) and inductance (Li) of each apparatus (other than the terminators) connected to
the Fieldbus must be less than or equal to:
FISCO: 5 nF and 10 H.
FNICO: 5 nF and 20 H
Revision date:
Version /Revision
2009-09-29
V3/R0
5350V111
5350QE01
WWW.PRELECTRONICS.COM
), and maximum power Pi (Pmax),
MAX
) and current (Io or I
or I
) and the power Po
t
SC
t
) for each intrinsically device
i
) which can be
a
) for each intrinsically device and
i
) which can be safely
a
or I
, and C
and L
for barriers are
SC
t
a
a
Page:
4/6