Installation; Setting Up; Service Unit - Amann Girrbach Ceramill motion Modo De Empleo

Ocultar thumbs Ver también para Ceramill motion:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

I N S T A L L A T I O N
5

Installation

E N
5.1

Setting Up

_ The machine is intended exclusively for use
within dry, closed rooms.
_ Minimum space requirements for the Ceramill
Motion (incl. connections):
▪ Width: 615 mm
▪ Depth: 430 mm
▪ Height: 730 mm
_ The set-up surface must withstand a load of
62 kg.
_ During operation, the room temperature must
be between 18 °C and 30 °C; avoid large varia-
tions in temperature.
▷ Set up the machine on a sturdy work table or
work bench (off the ground and clear of walls).
▷ Adjust the height of the machine by screwing
the feet in or out, ensuring that the machine is
positioned firmly on all four feet.
Countries with 100-115 V mains voltage
When the mains voltage at the set-up location is
100-115 V:
▷ Pull out the fuse compartment ([11] in Fig. 2 on
page 7) below the mains plug.
▷ Replace both T3.15 A fuses with the provided
T6.3 A fuses.
30
5.2

Service Unit

The Jäger spindle is equipped with a sealing-air
feature. This sealing air prevents shavings/chips
and dust from entering the spindle. The service
unit filters possible contamination in the sealing
air and thus prevents costly machine damage.

Publicidad

Tabla de contenido
loading

Tabla de contenido