Table 2: Guidance and Manufacturer's Declaration: Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the devices should assure that it is used in
such an environment.
Immunity Test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC
61000-4-4
Surge
IEC
61000-4-5
IPL3500_18817_V2.indd 55
IEC 60601
Compliance
Test Level
Level
± 8 kV contact
± 8 kV contact
± 15 kV air
± 15 kV air
± 2 kV for power
± 2 kV for power
supply lines
supply lines
± 1 kV for input/
NA
output lines
2 kV
2 kV
0, 90, 180
0, 90, 180
& 270
& 270
degree phase
degree phase
angles
angles
55
Electromagnetic
Environment Guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.