Table 3: Guidance and Manufacturer's Declaration: Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in such
an environment.
Immunity
Test
Conducted
RF
IEC
61000-4-6
Radiated
RF
IEC
61000-4-3
IPL3500_18817_V2.indd 57
IEC
Compliance
60601-1-2
Level
Test Level
6 V/rms
6 V/rms
150 kHz to
150 kHz to
80 MHz
80 MHz
10 V/m
10 V/m
80 MHz
80 MHz
to 2,7 GHz
to 2,7 GHz
Electromagnetic Environment
Guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the device, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter. Recommended separation
distance.
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, should be
less than the compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked with
the following symbol: