5.3.4 Selecting a liquid type

The liquid type determines the limits of the set temperature range.
Selection of the liquid types below changes the settable temperature range as follows:
Liquid
Set temperature range
Water
0°C to 100°C
Water-Glycol
*-30°C to 70°C
Low Temp Oil
-50°C to 50°C
70°C to 150°C (TX150)
High Temp Oil
¥
70°C to 200°C (TXF200)
§-50°C to 150°C (TX150)
Override
§-50°C to 200°C (TXF200)
* Water-Glycol (50% water, 50% antifreeze (inhibited ethylene glycol)
High Temp Oil (silicone fluid with the following characteristics: viscosity 20 centistokes,
flash point ≥230°C, fire point ≥280°C).
¥
High Temp Oil (silicone fluid with the following characteristics: viscosity 50 centistokes,
flash point ≥285°C, fire point ≥340°C).
§
When set to Override it is the user's responsibility to select a suitable safe liquid.
See section 4.6 for full description of recommended bath liquids.


.
°c




:
.
Settings
Programs
Liquid
Probe
Alarms
Settings
Programs
Liquid
Probe
Alarms
30424 V6
Page 25
1. Rotate the dial until the settings icon is
highlighted, press the S button.
°c
2. Rotate the dial to scroll up and down the list until
°c


.
liquid is highlighted, press S to select.
Liquid options are displayed.
3. Rotate the dial to scroll through the liquid options,
°c


.
press S to save the selection.
The display returns to the settings screen.
Water
4. Press F to return to the home screen.
www.grantinstruments.com
TX150 & TXF200
Operating Manual