Troubleshooting - Grant Optima TX150 Manual Del Operador

Ocultar thumbs Ver también para Optima TX150:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

11.0 Troubleshooting

Symptom
Possible cause
Temp continues
Set temp is higher
to rise when not
than liquid temp
expected
Set temperature
Liquid type set does
too restricted
not allow required set
point
Temperature
Set temp is lower
does not rise
than liquid temp
when expected
Set temperature is
too close to ambient
Display shows
High temperature
"High
warning alarm has
Temperature
tripped
Alarm"
Display shows
Low temperature
"Low
warning alarm has
Temperature
tripped
Alarm"
Unit showing
Calibration values not
erratic
set correctly
temperatures
New calibration
Incorrect calibration
point is not
value
saved
Only the
Unit is in standby
Standby icon
mode
can be
highlighted
Icons can be
Program running
highlighted but
not selected
30424 V6
Page 39
Action required
Check that the bath set temperature is correct (see
section 5.2.9).
Change to different liquid type (see section 5.3.4).
Check that the bath set temperature is correct (see
section 5.2.9).
Increase the set temperature (see section 5.2.9) or
fit accessory cooling (see section 4.9).
Check that the bath set temperature is correct (see
section 5.2.9).
Check that high temperature alarm is correct
(section 5.3.6).
Check that the liquid level in the bath is adequate
(see section 6.4 for minimum fill levels).
Check that the bath set temperature is correct (see
section 5.2.9).
Check that low temperature alarm is correct (section
5.3.6).
Check that the liquid level in the bath is adequate
(see section 6.4 for minimum fill levels).
Restore the factory calibration settings (see section
5.4.1) then re-calibrate if required (see section 5.4).
The temperature calibration point is less than 20°C
from an existing calibration point – choose a higher
temperature (see section 5.4).
The measured liquid temperature is more than 5°C
away from selected calibration temperature.
Highlight and Select the Standby icon and Press "S"
to exit Standby mode.
Wait until program has finished or stop program.
www.grantinstruments.com
TX150 & TXF200
Operating Manual

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Optima tx150lOptima txf200Optima txf200l