7. What do I do if ...
... no weight is displayed when there is a load on the scale?
– Is the scale switched on?
– Is the ON-OFF switch in position "ON"?
– Check the batteries.
... the scale keeps switching on, while being transported, for example?
– The start key has been pressed. Due to the very low energy requirement of the scale,
the service life of the batteries will not be impaired. For longer journeys, you can push
the ON-OFF switch in position "OFF" if necessary.
... the scale displays a weight after being transported or after new batteries have
been put in?
– Press the Start key again. The scale will then work normally again.
0. 0 0
...
does not appear before weighing?
– Press the Start key again - there should not be any load on the scale.
-----
...
appears instead of
– Press the Start key again - there should not be any load on the scale.
one segment is illuminated constantly or not at all?
– The relevant segment has a fault. Inform Servicing.
the display shows
– Battery voltage is running low. Change the batteries in the next few days.
bAtt
appears in the display?
– Batteries are empty. Put in new batteries.
St0p
appears in the display?
– Maximum load has been exceeded.
the display flashes?
– Take the load off the scale and start the scale again. Wait until
weigh again.
Er
... the display
– Start the scale again after it switches off automatically. The scale will then work nor-
mally again.
If this is not the case, disconnect the power supply by briefly switching off the scale at
the ON-OFF switch. If this measure is equally unsuccessful, inform the Service depart-
ment.