3-Phase 3-Wire (3 3W) Unbalanced-Load Power Function - Amprobe ACD-45PQ Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

• After the kWHr Recording session is stopped ("
result will supersede the previous one stored in the non-volatile memory.
You can then switch off the meter for transportation, storage, or even
battery changing with memory remained.
• To avoid mis-storage to memory, it is important to properly stop ("
kWHr Recording session before sliding the slide-switch function-selector
to any other function positions.
3-Phase 3-Wire (3
• Set the slide-switch function-selector to the "3~Un-Bal" Power position.
Press "3W•4W" button momentarily to select 3-Wire measurements.
Annunciator "3W" turns on.
• Clamp the jaws around "Line 1" as reminded by annunciators "
and connect Black test probe (COM terminal) to "Line 3" and Red test
probe (+ terminal) to "Line 1" as reminded by annunciators "
mini-display.
• When the reading is stable, press "
first measuring value.
• Then clamp the jaws around "Line 2" as reminded by annunciators "
L2", and connect Black test probe (COM terminal) to "Line 3" and Red
test probe (+ terminal) to "Line 2" as reminded by annunciators "
" on mini-display.
• When the reading is stable, press "
second measuring value. The meter will then calculate, store and display
the total 3-Phase Power result automatically. Annunciators "
turn on.
• Press "
" button momentarily again for new measurements.
• Press "SELECT" button momentarily to view ("
result. Annunciator "
momentarily again to continue ("
3W) Unbalanced-Load Power Function
" turns on & flashes. Press "SELECT" button
" button momentarily to enter the
" button momentarily to enter the
").
11
") properly, the new
L1 L2 L3"
") the last stored
")
L1",
" on

Publicidad

Tabla de contenido
loading

Tabla de contenido