Remote Control handling –
Patients should avoid
• Immersing the Remote Control in liquid as this could damage the device.
• Dropping the device or mishandling it in any way that may damage it.
Patients should clean the device with damp soft cloth.
Remote Control use –
Patients should avoid operating the Remote Control when near flammable or explosive gases.
Storage and Usage Environment
Component packaging –Do not use the component if any of the following have occurred:
• The storage package has been damaged, pierced, or altered.
• The component itself shows any signs of damage. The component may not function properly.
• The use-by date has expired. In this case, component performance cannot be guaranteed.

Usage environment:

The following lists the appropriate temperature, humidity, and pressure usage conditions for use of the TS:
• Temperature: 5 °C to 40 °C
• Humidity: 15% to 95%
• Pressure: 106 kPa to 70 kPa
Shipping and Storage environment:
The following lists the appropriate temperature, humidity, and pressure conditions for shipping and storing the TS:
• Temperature (short term: 3 days): -25
C to 70
C
o
o
• Temperature (long term): 20
C to 30
C
o
o
• Humidity (short term: 3 days): 15% to 95%
• Humidity (long term): 30% to 85%
• Pressure (short term: 3 days): 57 kPa to 106 kPa
• Pressure (long term): 70 kPa to 106 kPa
If the TS was stored at temperatures outside of this range, do not use it until it has returned to the operating temperature range.