• If there is no flashing light when you press the button, try pressing the button again. Confirm you are pressing the button as
shown in Figure 2. If there is no light after multiple attempts at pressing the button, do not use this TS. Activate a new TS. If the
issue persists with a new TS, contact Axonics.
Connecting the Trial Stimulator to the Tined Lead or PNE Lead
The following section describes the process for connecting the TS to the cables used to connect to the tined lead or the PNE lead. This
should be performed when an Axonics lead has already been implanted and the appropriate cables are available. For a trial using
the PNE lead, the Basic Trial Cable should be connected to the TS. For a trial using the tined lead, a Percutaneous Extension (PE) cable
should be connected to the TS.
1. Align the raised base on the TS connector with the raised bar on the Basic Trial Cable or the grey colored bar on the PE (Figure 3).
2. Press the connectors together to connect the Basic Trial Cable or PE to the extension of the TS (Figure 3).
3. Use the CP to check the TS impedances to verify the TS and cable is connected correctly.
Notes:
• There will not be any gap between the connectors when fully inserted.
• Use the CP to check the electrode impedances to confirm that the cables are connected.
Figure 2: Activate the TS by pressing the button on the back of the stimulator
15