ARTROMOT S3 Manual Del Usaurio página 53

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 124
– It is not permitted to change the
adjustment of 1 through 7 while a
patient is sitting in the CPM device.
– Movements must not cause any
pain or irritation.
– Patients must be fully conscious
while being instructed in the use of
the CPM device and during therapy.
– The choice of the therapy parame-
ters to program and of the therapy
protocols to use is restricted to
the responsible physician or
therapist.
It is the physician's or therapist's
decision whether or not to use the
CPM device on a specific patient
– The patient must be familiar with the
functions of the ARTROMOT
programming unit and the unit
must be within easy reach of the
patient, allowing him or her to stop
therapy, if needed. Patients unable
to operate the programming unit,
e. .g. paralytic patients, must never
be left unattended during therapy.
– Write the patient's name on the
patient chip card. The card should
only be used for this patient. If the
patient chip card is used for another
patient, be sure to delete the
previous patient's data from the
card first (see sections 4.1 and 5.3,
paragraph "New Patient").
Use original chip cards only.
– All accessories used with the
ARTROMOT
®
-S3 device must first
be approved by ORMED.
– Do not allow parts of the body or
any objects (such as blankets,
cushions or cables) to get caught in
the moving parts of the CPM
device.
Warning!
Shock hazard —
Strictly observe the following war-
nings. Failure to do so endangers the
lives of the patient, the user and other
persons involved.
– Before use allow the
ARTROMOT
temperature. If the device has been
transported at temperatures below
0 °C (32 °F), leave it to dry at room
temperature for about 2 hours, until
any condensation has disappeared.
– The ARTROMOT
operated in dry rooms.
– When disconnecting the device from
the power line, remove the plug
®
-S3
from the wall outlet first, before
disconnecting the cable from the
device.
– When connecting the device to
other equipment or when creating a
medical system, check that the sum
of leakage currents will not cause
any hazard. Please contact ORMED,
if you have questions in this matter.
– Do not use multiple portable socket
outlets (MPSO) to connect the
device to the power line.
ARTROMOT
ted to a properly installed wall
outlet with a non-fused earthed
wire. Before connecting the power
cord, it must be completely unrolled
and placed such that it will not get
caught in the moving parts of the
device.
– Before cleaning and service inter-
ventions, disconnect the device
from the power line by removing
the power cord from the wall outlet.
– Liquids must not be allowed to
enter the CPM device or the
programming unit. If liquids have
entered into the devices,
ARTROMOT
ately checked by a service techni-
cian, before it can be reused.
53
®
-S3 to reach room
®
-S3 must only be
®
-S3 must be connec-
®
-S3 must be immedi-

Publicidad

Tabla de contenido
loading

Productos relacionados para ARTROMOT S3

Tabla de contenido