and then to the programmed internal
rotation position.
Subsequently the carriage reverses
5° towards external rotation and then
moves very slowly back again to the
programmed internal rotation position
(display
). After that it attempts to
attain another 5°, moving even slower
than before (display
If a high resistance toward the
additional 5° is sensed, the reverse
on load function is automatically
activated and the carriage moves in
the opposite direction.
This stretch cycle is repeated ten
times.
After that the carriage moves to the
programmed maximum external
rotation position and restarts the
internal rotation stretch cycle.
This sequence can be repeated as
often as needed. You can stop
treatment with the STOP key; after
the programmed therapy duration, it
will stop automatically.
Default: disabled
Note!
• If a pause is programmed when you
enable the function, the carriage
will stop for the duration of the
pause each time the maximum
stretching value is attained.
• Enabling the stretching in internal
rotation function automatically
disables motor A.
The only movement performed is
internal / external rotation.
• It is not possible to combine a
second stretching function with the
special function "stretching in
internal rotation".
• During operation, the display will
show the symbol
upper, left-hand
corner.
I Stretching in external rotation
With the special "Stretching in
external rotation" function the car-
).
• If a pause is programmed when you
• Enabling the stretching in external
in the
• It is not possible to combine a
• During operation, the display will
69
riage gently stretches the joint
towards the outside.
Starting at the middle position the
carriage will first move to the pro-
grammed internal rotation position
and then to the programmed external
rotation position.
Subsequently the carriage reverses
5° towards internal rotation and then
moves very slowly back again to the
programmed external rotation posi-
tion (display
). After that it
attempts to attain another 5°, moving
even slower than before (display
).
If a high resistance toward the
additional 5 ° is sensed, the reverse
on load function is automatically
activated and the carriage moves in
the opposite direction.
This stretch cycle is repeated ten
times.
After that the carriage moves to the
programmed maximum internal -
rotation position and restarts the
external stretch cycle.
This sequence can be repeated as
often as needed. You can stop
treatment with the STOP key; after
the programmed therapy duration, it
will stop automatically.
Default: disabled
Note!
enable the function, the carriage
will stop for the duration of the
pause each time the maximum
stretching value is attained.
rotation function automatically
disables motor A. The only move-
ment performed is internal / external
rotation.
second stretching function with the
special function "stretching in
external rotation".
show the symbol
upper, left-hand
corner.
in the