Problem/Possible cause
You may have removed the
SD card before settings were
copied to the device.
The following message is NOT displayed on the LCD after you try to update the settings using
Settings updated successfully, press any key
the SD card:
The settings were not updated.
Water splashing on your face from the H5i
The water tub is overfilled.
Condensation is forming in the
air tubing and mask.
Leaking water tub
The water tub may be damaged
or cracked.
The cleanable water tub is not
assembled correctly.
Air feels too warm/cold in the mask
The temperature of the
ClimateLine or ClimateLine
heated air tubing is set too
high/low.
Solution
Reinsert the SD card and wait for the Home screen or the
"Settings updated successfully, press any key"
appear on the LCD.
Note: This message only appears once. If you re-insert the SD card
after you have updated your settings, the message will not be re-
displayed.
Contact your clinician/service provider immediately.
Check that the water level is below the maximum water level
mark.
Turn the humidity level setting down via the menu.
Contact your service provider for a replacement.
Check for damage and reassemble the cleanable water tub
correctly.
Turn up/down the heated air tubing temperature via the menu.
MAX
message to
English
7