Humidifier Performance; Pneumatic Flow Path - ResMed S9 Serie Guia De Informacion

Ocultar thumbs Ver también para S9 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

When using the SlimLine or ClimateLine above 20 cm H
may not be reached if used with an antibacterial filter. The device performance must be checked
prior to prescribing the SlimLine for use with an antibacterial filter.
The ClimateLine or ClimateLine

Humidifier performance

The following settings have been tested at 71.6°F (22°C) ambient temperature:
Mask pressure
cm H
O
2
3
10
20
25
a. AH - Absolute Humidity in mg/L.
b. BTPS - Body Temperature Pressure Saturated.

Pneumatic flow path

MAX
is designed only for use with the H5i.
RH output %
Setting 3
90
95
95
100
O, the device optimum performance
2
Nominal system output AH
Setting 6
100
100
100
100
1. Flow sensor
2. Blower
3. Pressure sensor
4. Mask
5. Air tubing
6. H5i
7. Device
8. Inlet filter
Setting 3
10
11.5
11
12
, BTPS
a
b
Setting 6
18
21
18
13.5
English
11