5.3 Cleaning CLICK 2

3. Then rinse the parts in clean water.
4. Vigorously shake the lower part to remove any water that may have entered
the heater rod contact. Dry the parts thoroughly with a soft cloth. The heater
rod contact must be dry.
EN
28
Hygienic preparation
not
Do
touch the heater rod. The metal takes a long
time to cool down.
1. Open the humidifier by twisting the upper and
lower parts against each other in an counter-
clockwise direction.
2. No water must penetrate into the pressure line
during cleaning or this could lead to incorrect
pressure readings. Therefore the pressure line
must remain sealed.
To seal the pressure line at the outlet-
connections push the silicone tube of the closure
(supplied as standard CLICK 2) over the end of
the pressure line.
To seal the pressure line at the inlet connections,
push the thin end of the closure by a slight rota-
tional movement into the seal of the pressure
line.
Clean the upper and lower parts with lukewarm
water to which a little detergent has been added.
We recommend cleaning the housing parts once
a week in the upper tray of the dishwasher at a
maximum temperature of 65 °C.