Seca 763 Manual Del Usuario página 26

Ocultar thumbs Ver también para 763:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 58
... the head stop slides down and does not stay in the required position?
– Place the scale carefully on its side so
that the base of the scale is accessible.
Insert the Allen key supplied in the rec-
tangular hole on the end of the measu-
ring rod as shown in the illustration and
tighten up the screw for the measuring
rod guide clockwise.
The measuring rod with the head stop
should remain freely mobile, but not be
able to slide down on its own.
Then stand the scale upright again.
8. Servicing
Your seca scale leaves the factory with an accuracy greater than ± 0,15 %. To ensure that
this accuracy is maintained, the product must be carefully set up and regularly maintained.
Depending on the frequency of use, we recommend servicing every 3 to 5 years. Please
ensure that a qualified service agent is used. If in doubt, contact seca Customer Service for
further information.
26

Publicidad

Tabla de contenido
loading

Tabla de contenido