KPA 02
Digital clamp meter
Technical specifications
Environment temperature: 23±5°C, relative humidity (RH): <75%
DC Voltage
Range
660mV
6.6V
66V
660V
1000V
Input impedance: 10MΩ
Overload protection: 660mV range: 250V DC or AC (RMS),
6.6V-1000V ranges: 1000V DC or 600V AC (RMS)
Maximum input voltage: 1000V DC
Note:
In the small voltage measuring range, when the probe is not connected with
the circuit to be tested, and the meter may have fluctuating readings, which
is normal and caused by the meter's high sensitivity. When the meter is
connected with the circuit to be tested, you will get actual measured value.
AC voltage
Range
660mV
6.6V
66V
660V
750V
Input impedance: 10MΩ
Overload protection: 660mV range: 250V DC or AC (RMS),
6.6V-750V range: 1000V DC or 750V AC (RMS)
Maximum input voltage: 750V AC (RMS)
Frequency range: 40 ~ 600Hz
Response: true RMS
Instructions manual |
Resolution
0.1mV
0.001V
0.01V
0.1V
1V
Resolution
0.1mV
0.001V
0.01V
0.1V
1V
26
Accuracy
±(0.8% of reading + 2 digits)
±(0.7% of reading + 2 digits)
±(0.8% of reading + 2 digits)
Accuracy
±(1.5% of reading + 5 digits)
±(0.8% of reading + 3 digits)
±(1.0% of reading + 4 digits)
www.grupotemper.com