Minebea Intec PR6130/65S Manual De Instalación página 58

Caja de conexiones de cable
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Hazardous location Class I, II, III, Division 2
ambient temperature: -20°C to +60°C
10)
Load cell #1
Rmin = 300W
10)
Load cell #2
Rmin = 300W
Load cell #8
Rmin = 300W
Hazardous location Class I, II, III, Division 1
ambient temperature: -20°C to +60°C
9)
Load cell
#1
9)
Load cell
#2
Load cell
#8
Datum
Name
Date
Erstellt
29.07.20
Hiller
Written by
Geprüft
03.08.20
Schallhorn
Reviewed by
Freigabe
Schallhorn
03.08.20
Released by
Output parameters:
The output voltage is
given by the connected
evaluation unit.
10)
Equipotential bonding conductor
Class I, Zone 0 or Zone 1 or
Zone 20 or Zone 21
9)
See page 2
Equipotential bonding conductor
Material
.
Ausgabe / Revision
Änderung / Alteration
03
PA50200657
Cable junction
box
6)
R1
6)
R2
Input parameters:
Um = Umax = 25V
6)
R8
8)
or
Cable junction
box
6)
R1
6)
R2
6)
R8
8)
Ben ennun g / Title
Control Drawing (PR6130/6... and PR6021/6...)
Zeichnung s-Nr. / Drawing number
36931-751-07
Evalaution
unit
non
10)
with
intrinsically
safe
output
to the load
cells
Evalaution
unit
9)
with
intrinsically
safe
output
to the load
cells
Maßstab /
Scale
1:1
Blatt
Sheet
Teildok. Nr./
von
Par t doc. no.
592
of
1
4

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Pr6130/68sPr 6130Pr 64sa