Specification of the SpiroUSB

General
Power supply:
Dimensions:
Weight:
Operating Temperature:
Operating Humidity:
Storage & Transport
temperature:
Storage & Transport
humidity:
Predicted Values:
Transducer:
Resolution:
Accuracy:
Measurements:
Input 5V 0.1A
50x60x90mm
130g
0 to +40° C
30% to 90% RH
-20 to + 70° C
10% to 90% RH
Various depends on National preference
Carefusion Bi-Directional Digital Volume.
10ml volume 0.03l/s flow
+/-3%. To ATS recommendations –
Standardisation of Spirometry 1994
update for flows and volumes.
See SPCS operating manual for details of
all the spirometry measurements
available.