Trouble Shooting Information
Should you encounter problems operating your Spiro USB meter,
please consult the table below:
Problem
Does not register a
blow
Blows are inverted on
the display
Blows tracing ends
abruptly although
patient still exhaling
Calibration procedure
failed or cannot be
completed
Possible Cause
Head assembly or
cable broken
Head assembly or
cable broken
Tubine sticking
Turbine may be
faulty
Turbine not
fitted tightly
to calibration
syringe
Calibration
syringe does not
have an
inspiratory seal
or seal is leaking.
Shaft of the
syringe being
pushed down
52
Solution
Replacement of head
assembly or return the
unit for servicing
Replacement of head
assembly or return the
unit for servicing
Clean turbine in warm
soapy water or sterilizing
solution, if problem
continues a replacement
turbine may be required
Repeat calibration
procedure, if problem
still persists, replace
turbine or return unit for
servicing
Ensure the syringe is
fitted to the turbine using
the adapter supplied part
# 861427. (supplied
with the syringe)
Ensure you are using
manufacturer
recommended syringe.
The syringe should be
emptied and filled with
one smooth stroke, avoid
pushing down on the
shaft or banging at the
end of each maneuver.