Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The SpiroUSB is intended for use in the electromagnetic environment specified below.
The customer or the user of the SpiroUSB should assure that it is used in such an
environment
Emission Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations /
flicker emissions
IEC61000-3-3
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The SpiroUSB is intended for use in the electromagnetic environment specified below.
It is the responsibility of the customer or the user of the SpiroUSB should ensure that
it is used in such an environment
Immunity Test
Electrostatic
discharge (ESD)
IEC61000-4-2
Electrical fast
transient / burst
IEC61000-4-4
Surge
IEC61000-4-5
Compliance
Group 1
Group B
Not Applicable
(USB powered)
Not Applicable
(USB powered)
IEC 60601 test
level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input /
output lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Electromagnetic Environment -
Guidance
The SpiroUSB uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment
The SpiroUSB is suitable for use in
all establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes
Compliance
Electromagnetic
Level
Environment -Guidance
± 6 kV contact
Floors should be wood,
± 8 kV air
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Supply line
Mains power quality should
tests not
be that of a typical
applicable
commercial or hospital
environment.
± 1 kV for
USB lead
Not Applicable
Mains power quality should
(USB
be that of a typical
Powered)
commercial or hospital
environment.
56