Periodical Maintenance Checks - Gima TOBI Manual De Istrucciones

Ocultar thumbs Ver también para TOBI:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

The TOBI MANUALE suction equipment does not need maintenance or lubrication.
It is necessary to check functioning and instrument before every use.
Unpack the instrument and always check integrity of plastic parts, they might have been damaged during previous use.
Close the aspirator outlet with your finger.
Press the foot and check vacuum functionally.
Type of fault
1. No aspiration
2. No aspiration
3. The Vacuum power on the
patient side is either very low or
absent
4. The float doesn't close
5. The float doesn't close
6. Low suction
Faults 1 - 2 - 3 - 4 - 5 - 6
If the overfill security system fails, continued aspiration will be stopped by the antibacterial filter.
Should this back-up security system also fail, there is a risk of penetration of liquid into the device.
Don't attempt to proceed with aspiration under these circumstances. Return the device to GIMA for service.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT
GIMA TECHNICAL SERVICE.
GIMA DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL SERVICE CHECKING, APPEARS
TO BE TAMPERED.

PERIODICAL MAINTENANCE CHECKS

Cause
Jar Cap not screwed on correctly
Lid seal not in its seat
• Protection filter blocked or
damaged
• Connection tubes blocked,
kinked or disconnected
• Shut-off valve blocked or
damaged
• Pump motor damaged
If the cap has been washed,
ensure that the float is not
partially detached
The float it's covered by dirty
material
Foam inside the jar
None of the procedures have
achieved the desired results
11
Remedy
Unscrew the cap, and re-screw it
Unscrew the cap and insert the seal properly in
its seat
• Replace the filter
• Replace or reconnect the tubes, check the jar
connections
• Empty the jar, or disconnect the tube from
the jar and unblock the shut-off valve.
The unit twill only work in the upright position
• Refer to authorised service personnel
Fit the float into it's place
Unscrewed the cap, leave the and put in on
autoclave
Fill the jar to 1/3 full of ordinary water
Contact GIMA customer service

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

28220

Tabla de contenido