Connect the short silicon tube 1, with antibacterial filter 2, to the suction connector 3.
The other tube, with one end connected to the filter must be connected with the other end to the jar's lid connector 4
where has been fixed the red float. When the 90% of the volume of the jar is reached there is the activation of the security
float(the float close the aspiration connector on the jar) to avoid liquid penetration inside the device.
The device must be used on a plan of horizontal operation.
WARNING: Ensure that the FLUID SIDE or IN marker on the filter is on the side facing the collection jar lid and fitted into the
"VACUUM". A wrong connection causes immediate destruction in case of contact with sucked liquids.
Connect the long silicon tube 5 to the other jar's lid connector 6
Connect the other end of the long silicon tube 5 to the probe plastic connector 7 then connect the suction probe 8 to it.
Unscrew the jar's lid and fill the jar 1/3 full or ordinary water (this for an easy cleaning operations and an rapid reaching
of the functionality vacuum) the rescrew the lid on the jar correctly.
Push foot 9 to start suction.
To extract the accessories and start with cleaning as write in the clean chapter
MAKE SURE THAT CHILDREN AND/OR MENTALLY ILL PEOPLE DO NOT USE THE DEVICE WITHOUT ADULT
INSTRUCTION FOR USE
NEVER USE THE DEVICE WITHOUT JAR AND / OR PROTECTION FILTER
SURVEILLANCE