Transcend 3 miniCPAP Quick Guide
Using the Transcend 3 miniCPAP
The control panel of the Transcend 3 miniCPAP has two pushbuttons that activate the
blower and initiate the pressure ramp feature. There are also two LED lights, including a
green LED for indicating normal operational modes and a yellow LED for indicating fault
conditions. The Transcend 3 miniCPAP operational status is displayed by LED illumination
states.
When a power source is connected to the device the Transcend 3 miniCPAP power-up LED
flash sequence should initiate. Once the power-up LED flash sequence is complete the LED
lights will turn off. This sequence indicates that power is being supplied to the Transcend
3 miniCPAP and that it has successfully entered Standby Mode. During therapy delivery,
the LED lights remain off to avoid disturbing the patient and/or bed partner.
If the Transcend 3 miniCPAP loses power while delivering therapy it will resume
Note
delivering therapy as soon as power is restored and you press the power button.
The device will repeat the power-up LED flash sequence prior to the blower
restarting.
Standard user modes
Normal operation consists of four modes:
Off
When the device is not connected to a power source the device is off. Control
panel LEDs are both off.
Standby
When power is applied to the device it completes the power-up LED
sequence and enters Standby Mode. Standby Mode is also initiated by
pressing the power button when the device is in On Mode or if the mask is
removed while in On Mode. As long as power is supplied to the device it will
remain in Standby Mode until On or Drying Mode is initiated.
On
When in On Mode the blower is working and regulated device therapy
pressure is being generated. The LED lights remain off. On Mode is initiated
by pressing the power button when the device is in Standby Mode and the
mask is worn by the patient.
Drying
To initiate Drying Mode depress the Ramp button and press the Power button
simultaneously. When in Drying Mode the blower runs at a low speed for 30
minutes. During Drying Mode the LED lights remain off and blower pressure
is not regulated to provide therapy.