How To Test Control Solution
Use By Dates
LOT - ABC1234
EXP - 2018/10/31
1 - 72-90 mg/dL
Control Solution/Solución de Control
Level/Nivel 2
3 ml
2 - 150-200 mg/dL
Manufactured for/Fabricado para:
3 - 301-398 mg/dL
Fort Lauderdale FL 33309 USA
May 30, 2018
For in vitro diagnostic use only.
Store in a dry place at room temperature between
59-86F (15-30C). DO NOT REFRIGERATE
OR FREEZE.
U5TVH02 Rev. 40 - U5TVH72
Test Strip Label
Control Solution Label
(Examples only and do not represent
actual Control Test ranges)
1. Gather and check supplies.
See Getting Started -Testing Checklist.
Do not use test strips or control
solution if either Use By Dates (written
or printed) have passed. Discard and
use new products.
2. Allow control solution,
3. Wash hands.
vial of test strips and
Dry thoroughly.
meter to adjust to room
temperature.
18
5. Remove one
4. Gently swirl or
invert control
test strip from
solution bottle
vial. Close vial
immediately.
to mix.
DO NOT SHAKE.
Use test strip
quickly after
taking it out of
the vial.
7. Remove cap from
control solution
bottle. Gently
squeeze a drop
6. Insert test strip firmly
onto a clean tissue.
into Test Port. Meter
Wipe off bottle tip
turns on.
and discard tissue.
Gently squeeze a
Keep test strip in
drop onto a small
meter until testing is
piece of unused
finished. Do not add
aluminum foil,
control solution to test
clear plastic wrap,
strip before inserting
or waxed paper for
into meter.
testing.
19