Caution!
• Other disinfectants have not been tested. The effect of other disinfectants used interchangeably has not
been tested with the lancing device. Use of disinfectants other than Super Sani-Cloth Wipes may damage
lancing device.
Do not use the lancing device and call Customer Care for assistance toll free: 1-800-803-6025
or local: 1-954-677-4599 if:
~ Markings on lancing device are coming off or missing,
~ Trigger Button is hard to push or does not work,
~ End Cap does not go back on lancing device,
~ Arming Barrel does not click when gently pulled back.
~ Super Sani-Cloth Wipes have been tested on the lancing device for a total of 1095 cleaning and
disinfecting cycles, which is equal to cleaning and disinfecting the lancing device once a day for a 3 year
period.
50
Changing Battery
Push up to open
1. Turn meter
2. Turn meter over, tap gently on the
over. Lift tab on
palm of your other hand to loosen
Battery Door.
and remove the battery. Discard old
battery into an appropriate container.
3. Insert a new battery,
positive ("+") side facing
up.
~ Use non-rechargeable 3V
lithium battery, #CR2032.
51